Amprobe WT-30 Kullanıcı Kılavuzu - Sayfa 15

Ölçüm Cihazları Amprobe WT-30 için çevrimiçi göz atın veya pdf Kullanıcı Kılavuzu indirin. Amprobe WT-30 17 sayfaları. Salinity meter

Amprobe WT-30 Kullanıcı Kılavuzu

TROUBLE SHOOTING

Power On But No Display
• Make sure you press power key more than 100 mS.
• Check the battery conditions and replace if
necessary
• Move batteries away for one minute and then
re-install.
Display Disappear
• Check whether the low battery icon is appeared
before the display is off. If yes, replace with new
batteries
Air Bubbles Adhere On Electrode
• Stir the electrode completely and better to dip
the electrode into solution at oblique angle. After
soaking the electrode for 15~30 minutes, inspect
the electrode carefully to make sure no bubbles
adhere.
• If air bubbles still exist, tap the bottom of the
container gently and stir the electrode to remove
the air bubbles. If above method are not working,
remove the electrode out of solution and blow at
the electrode to remove the air bubbles
Error Code
Parameter: Salinity
"- - -"
Meter is in manual range 1 however the
conductivity measured value is higher than
10.0ppt.
• Press "p" more than 2 seconds to change
the range to range 2 or to auto ranging.
10