Figure 2: Desiccant Reservoir Pulled From H300P
Unscrew the reservoir plug and remove the spring, dust filter and metal mesh.
Remove the used desiccant. The used desiccant can be regenerated by heating it above 395°F
(200°C) for one to two hours.
Fill the reservoir with a fresh desiccant. Do not overfill or underfill.
Replace the dust filter, followed by the metal mesh, followed by the spring.
Screw the reservoir plug until tight. The spring should compress when the plug is tight.
If the spring does not compress, there is not enough desiccant in the reservoir.
If the spring compresses too much, so that it is excessively difficult to screw the reservoir plug, there
is too much desiccant in the reservoir.
H300P Temperature Humidity Chamber
Page 6