2. Infusion line without drip chamber: the operator must therefore use the reverse purge air
drainage technique expelling all air including that contained in the infusion bag and in the
infusion line. If there is no drip chamber, the infused flow must be estimated or checked by
means of a suitable flow regulator.
the operator must maintain pressure on this bag in a backwards direction to prevent it from sliding
forwards when closing the device.
2.
To close: pivot the cover forwards until it enters
into contact with the bulge at the surface of the infusion bag.
Stand directly above the device with outstretched arms and
press down on the cover, placing the cushions of each thumb
on the rough areas located on either side of the cover.
A force of approximately 36 kg is required to pressurise the
device. Prior training is thus recommended to master how to
operate the device.
WARNING: once the movement has been initiated (once the initial resistance has been broken), the
pivoting movement applied to the cover will become easier and the force exerted must be reduced.
This action places pressure on the mobile plate. The resulting significant downwards movement can
be observed. The system closes and is therefore armed as soon as the closing clicks are heard. After
releasing the pressure, the operator must check that the two fastening hooks visible through the
transparent cover are engaged.
If engagement can only be observed on one side, the plate will adopt an inclined position in relation
to the device. In this event, push back down on the cover to release the engaged hook and repeat the
closing operation. This operation pressurises the infusion bag.
2.4
Inserting the infusion bag with the infusion
line and closing the device
1.
Insertion: With the cover fully open, position the
infusion bag on the plate and attach it to one of the fastening
hooks on the plate.
The purpose of these fasteners is to prevent the bag from
sliding forwards when closing the device. Two fasteners are
available according to the model of bag used. Their location is
designed to suit the main infusion bag models available on the
market.
If the fastener does not correspond to the bag model used,
7 / 16