Droper Field 1000 Kullanım Talimatları Kılavuzu - Sayfa 9
Tıbbi Ekipman Droper Field 1000 için çevrimiçi göz atın veya pdf Kullanım Talimatları Kılavuzu indirin. Droper Field 1000 16 sayfaları. Mechanical pressure infuser
Ayrıca Droper Field 1000 için: Kullanım Kılavuzu (15 sayfalar)
Ideally, the device should be positioned level with the patient's midaxillary line
to guarantee an approximate pressure of 100 mbar.
If placed at a higher level, and according to the atmospheric pressure conditions, the pressure shall be
approximately increased (in mbar) by the distance (in cm) separating this line from the device outlet
and conversely if it is placed at a lower level. In any event, this position must be taken into account
when adjusting the flow rate.
show the first line. This line is the widest and coloured green. The next two lines are red in colour. The
appearance of the peak of the pyramid indicates that the plate has reached its end point.
2.
Draining an infusion bag installed in the Droper Field using the pressure
generated by the latter. To achieve this, position the Droper Field
upright on its two rear supports. Begin to expel the air as described
above in point 1.a and 1.b according to the type of line and whether a
drip chamber is used.
Connect the infusion line to the patient as per the operating
instructions specific to the infusion line. Check the free
tubing of the infusion line and the absence of any
folds. Adjust to the desired flow rate using the roller
clamp (case 2.3 1-a) or a suitable flow regulator
(case 2.3 1-b and 2.3 2) and begin the infusion.
stage may be assessed by the triangular shape drawn on the
front surface of the mobile plate and containing different
coloured lines. The rising action of the plate during the infusion will
The Droper Field 1000 is designed for use in
emergency
situations
environments and does not provide for the electronic
or electrical monitoring of the infusion. The infusion
and
in
hostile
9 / 16