8. COMMISSIONING

8.1 Function check

Before using for the first time check set-up
and all cables carefully.
• Switch on power at the VistaCam2.
• Switch on lighting. Yellow LED shows.
• Image viewer appliance should be switched
on.
• The appliance is now operational and an
image should be seen in the image viewing
appliance.

8.2 Test

This appliance was checked before transport
that it satisfied all electrical safety
requirements. As part of the commissioning
procedure electrical testing should be carried
out and documented. Furthermore, electrical
checks should be carried out at regular
intervals in line with national regulations.
9. TRANSPORT OF THE
VISTACAM 2
If the VistaCam 2 is transported from one
treatment room to another the camera can be
uncoupled from the Docking-Station without
being switched off. After replacing in the
Docking-Station however, the camera must be
switched on again.

WASTE DISPOSAL

10. APPLIANCE DISPOSAL

The electronic board and components should
be disposed of as electronic waste. The other
elements can be disposed of as normal waste
observing local rules and regulations.