dymax BlueWave 50 Kullanım Kılavuzu - Sayfa 12

Aydınlatma Ekipmanları dymax BlueWave 50 için çevrimiçi göz atın veya pdf Kullanım Kılavuzu indirin. dymax BlueWave 50 20 sayfaları. Uv light-curing spot lamp system

OPERATION

®
The BlueWave
50 will arrive almost fully assembled. Please
refer to the Installation and System Interconnect section of this
manual for instructions on installing the Lightguide (sold
separately), Power Cord, and Footswitch.
The system should be positioned in a dry location that does not
obstruct airflow from the rear of the unit. To energize the system,
turn the Power Switch from the "O" position to the "l" position
(Figure 10). In the "I" position, the Fans, Hour Meter, Timer, and
50W Bulb should begin to function. This can be confirmed by
viewing the illuminated Bulb Status Indicator Light above the
Lightguide Mount (Figure 11). Before operating unit, allow the
50W Bulb to warm-up for approximately 5 minutes.
CAUTION:
Always wear protective goggles or face shield when
working near UV light. Never look directly at light exiting
Lightguide.
PRÉ-CAUTION:
Toujours porte lunettes de protection ou
protéger de visage en travaillant près lumière ultraviolette.
J'amais regarde directement à lumière sortie de le guide de
lumière.
The Timer located on the front panel of the BlueWave 50 is
factory set to the most common operating mode and
recommended operation of the Timer with the BlueWave. Some
modes available on the Timer may not operate correctly with the
BlueWave unit.
The front panel of the Timer contains a LCD Display and Keypad. The LCD Display has a Power Indicator, Signal
Input Indicator, Reset Indicator, Gate Input Indicator, Key Protect Indicator, Output Indicator, Preset Value, Set
Value, and Timing Operation Indicator. A brief description of each display and location:
• Power Indicator – Displays PW in upper left corner of Timer Display. Displayed when external power for
operation is present.
• Signal Input Indicator – Displays SIG in upper left corner of Timer Display, located to right of Power Indicator.
Displayed when activate signal is present.
• Output Indicator - Displays OUT in upper right corner of Timer Display. Displayed when Relay is switched to
NO; is not displayed when Relay is switched to NC.
• Present Value – Four digits segmented display in center of Timer. Shows current status of time.
• Set Value – Four digit segmented display in lower right corner of Timer. Shows set length of time.
• Timing Operation Indicator – Displays RUN in upper right corner to the left of Output Indicator. Displayed
when Timer is active.
To operate the Timer, select the "timed" option of the Switch on the front panel. Program the time into the Timer
and depress the Footswitch. Factory settings will open the Shutter and the present value will begin to count
backward. When the Timer reaches 00.00, it will reset the value to the set value and close the Shutter. The
Timer cannot be stopped once it's started. If the power is removed from unit, the Timer will reset to the set value.
To select the time, press the appropriate up key until the corresponding digit increments in the set value. By
pressing the up key labeled (1), it will increment the left most digit on the Set Display. By pressing the up key (2),
12
Figure 10. Power Switch
Figure 11. Bulb Status Indicator Light
®
BlueWave
50 Light-Curing Spot Lamp