Display
Block Charts
Power On
Measuring Status
ENT
Press
Pass Code
(P.Cod)
ENT
Press
P.Code Correct
NO
YES
System Setting
Press
AL
Page (SYS)
ENT
Press
Control DI 1
Setting (di-1)
ENT
Press
Control DI 2
Setting (di-2)
ENT
Press
Decimal Point 1
Setting (dP1)
ENT
Press
Display Low Scale 1
Setting (dSPL1)
ENT
Press
Display Hi Scale 1
Setting (dSPH1)
ENT
Press
2nd Display
NO
YES
Decimal Point 2
Setting (dP2)
ENT
Press
Display Low Scale 2
Setting (dSPL2)
ENT
Press
Display Hi Scale 2
Setting (dSPH2)
ENT
Press
Display Average
Setting (AvG)
ENT
Press
Display Low Cut
Settingn(LCUt)
ENT
Press
Pass Code
Setting (CodE)
ENT
Press
Key Lock
Setting (LoCK) but cannot modify any values.
ENT
Press
Display
Input signal is over 120% of input range.
Input signal is under -20% of input range.
Input signal is over 180% of input range or meter error.
**Please check the wiring connection is correct first, if the problem still exist, please return the meter to the factory.
Display
Measuring Status
ENT
Press
&
together for 3 sec
AL
Input Low Scale
Calibration (inLo)
ENT
Press
Input Hi Scale
Calibration (inHi)
ENT
Press
System Setting
Page (SYS)
Press
&
together for 3 sec
Warning: Calibration of this meter requires a standard signal with 0.01% accuracy or
better and an external meter with 0.005% accuracy or better.
ENGINEER LEVEL OPERATING PROCEDURES
Present value for measurement
to enter pass code.
Press
AL
Pass code is correct that will enter to parameter groups.
Pass code is wrong that will back to measuring status.
Press
to select control DI parameters (2DSP, Z, MAX, HD); terminal 5 & 7 shorts.
EX: D/I 1 is set for (AZ), when terminal 5 (DI1) & 7 (COM ) shorts, the reset indicator (Z)
will flash & the value will be reset 0. (Default: AZ: D/I 1 reset)
Press
to select control DI parameters (2DSP, Z, MAX, HD); terminal 6 & 7 shorts.
EX: D/I 1 is set for (2DISP), when terminal 6 (DI2) & 7 (COM ) shorts, the display value
2 indicator (DISP2) will flash & show 2nd value. (Default: MAX: D/I 2 Max value)
Press
to select decimal point (0, 1, 2, 3, 4)
AL
EX: if the value shows "0.00" that means the decimal point is 2 digits.
to modify display low scale for the input signal zero value.
Press
AL
EX: If the input signal is 4~20mA; 4mA is shown display 0.00,
this parameter must be set for 000.00.
to modify display high scale for the input signal span value.
Press
AL
EX: If the input signal is 4~20mA; 20mA is shown display 100.00,
this parameter must be set for 100.00.
The following steps are only available when DI1 or DI2 is set for "2nd Display".
Press
to select decimal point (0, 1, 2, 3, 4)
AL
EX: if the value shows "0.00" that means the decimal point is 2 digits.
to modify display low scale for the input signal zero value.
Press
AL
EX: If the input signal is 4~20mA; 4mA is shown display 0.00, this
parameter must be set for 000.00.
Press
to modify display high scale for the input signal span value.
AL
EX: If the input signal is 4~20mA; 20mA is shown display 100.00, this
parameter must be set for 100.00.
Press
to modify display average (1~99)
AL
PS: Please use this function for stable display value when input signal is unstable.
Press
to modify display low cut to 0(0~99)
AL
Press
to modify pass code(0~19999)
AL
PS: Please don't forget the new pass code after modification.
Press
to lock the keys, using key lock function only can view the parameters,
PS: no(unlock) ,YES("ENT" unlock , others lock)
Error Code of Self-Diagnosis
Descriptions
CALIBRATION OPERATING PROCEDURES
Present value for measurement
Press ENT &
AL
1. Input standard low scale signal.
2. Press
AL
1. Input standard hi scale signal.
2. Press
AL
1. Finish calibration operating procedures will enter to system setting group.
2. Press
&
Descriptions
Display
Input signal is over display range (99999)
Input signal is under display range (-19999)
EEPROM reading /writing suffers the interference (about 1 million times)
Descriptions
together for 3 sec will enter to calibration operating procedures.
to calibrate input low scale. (LED flashing)
to calibrate input hi scale. (LED flashing)
together to back to measuring status.
P2
Customers
specify
Customers
specify
Customers
specify
Customers
specify
Customers
specify
Customers
specify
Descriptions
Default
Default