V S A Q u i c k S t a r t G u i d e
the up and down arrows. Choosse any value between
0.003 and 1.0
a
the
values for equilibration and consequently the
w
resolution. If starting
a
0.10
w
0.3 ... up to the final
either equilibrium is achieved based on the trigger value
or it reaches the timeout value. Setting a higher
value makes the test faster, but decreases the quantity
of equilibtation points.
•
To choose the trigger %dm/dt value that indicates
equilibration, press enter with "Trigger" highlighted
and then cycle to the desired value using the up and
down arrows. The trigger can be shut off or be any
value between 0.001-1%/hr. Choose the number of
events required to meet the trigger value. The trigger
value represents an acceptable change in mass per
change in time to indicate equilibrium for a given step.
Setting a higher value will make the test faster, but
may not result in complete equilibrium. If the trigger
is turned off, the sample will be held at each step for
the time indicated by the timeout setting.
•
To choose to include a timeout setting, press with
"Timeout" highlighted and then cycle to the desired
value using the up and down arrows. You can select any
value between 5.0 min to 30 days. The timeout
determines the maximum time allotted for each
NOT the time from starting
value is off, which means there is no timeout value
and only the trigger value determines the end of step.
If you set both a trigger value and timeout value,
whichever VSA reaches first determines the end of step.
•
To choose to have the isotherm loop, press with
"Loop" highlighted to toggle between on and off.
Turning loop on causes the isotherm to automatically
return to the start
the final
both an adsorption and desorption curve (or vice versa
depending on the starting direction).
a
. The step setting for DVS determines
w
results in humidity being controlled to 0.1, 0.2,
a
a
(with the same settings) once
w
a
has been achieved. This would create
w
a
is 0.1 and has a step setting of
w
, remaining at each level until
w
a
to final
w
a
step
w
a
. The default
w