Decagon AQUA LAB VSA Hızlı Başlangıç Kılavuzu - Sayfa 11
Ölçüm Cihazları Decagon AQUA LAB VSA için çevrimiçi göz atın veya pdf Hızlı Başlangıç Kılavuzu indirin. Decagon AQUA LAB VSA 12 sayfaları. Vapor sorption analyzer
•
The delete entry is used to remove an unwanted stage.
• Follow the message on the screen to place an empty
stainless steel cup in the chamber and seal it. Press
tare the cup.
• Follow the message on the screen to place the sample in
the cup and press to weight the sample.
• Press to begin the test.
5. Running a water activity test—
isotherm test is running)
• Remove the weighing pan.
• Press until you are at the Measurement Tab (2nd tab).
Move latch left and/or press to start a reading. The
pinwheel will spin indicating a measurement is running.
a
• The
reading for the sample and the temperature will
w
display when finished.
• To change between dew point and capacitive readings,
you will need to toggle the Volatiles Yes/No option in
the Configuration Tab.
Volatiles = Yes g Capacitive Sensor
Volatiles = No g Dew Point Sensor
Note: Accuracy changes to +/- 0.02 when volatiles "Yes"
is selected.
V S A Q u i c k S t a r t G u i d e
(Can not be done when
to