Decagon AQUA LAB VSA Hızlı Başlangıç Kılavuzu - Sayfa 11

Ölçüm Cihazları Decagon AQUA LAB VSA için çevrimiçi göz atın veya pdf Hızlı Başlangıç Kılavuzu indirin. Decagon AQUA LAB VSA 12 sayfaları. Vapor sorption analyzer

The delete entry is used to remove an unwanted stage.
• Follow the message on the screen to place an empty
stainless steel cup in the chamber and seal it. Press 
tare the cup.
• Follow the message on the screen to place the sample in
the cup and press   to weight the sample.
• Press   to begin the test.
5. Running a water activity test—
isotherm test is running)
• Remove the weighing pan.
• Press   until you are at the Measurement Tab (2nd tab).
Move latch left and/or press   to start a reading. The
pinwheel will spin indicating a measurement is running.
a
• The 
 reading for the sample and the temperature will
w
display when finished.
• To change between dew point and capacitive readings,
you will need to toggle the Volatiles Yes/No option in
the Configuration Tab.
Volatiles = Yes g Capacitive Sensor
Volatiles = No g Dew Point Sensor
Note: Accuracy changes to +/- 0.02 when volatiles "Yes"
is selected.
V S A Q u i c k S t a r t G u i d e
(Can not be done when
to