14
Set a continuous current of 0.5A on output channel 3
RS3,0.5
10.1.3
Pulsed Output Mode
Set the pulse parameters.
RT
This command sets a channel to pulsed output. Note that output channel 1 can
only be triggered by INPUT1 and output channel 2 can only be triggered by
INPUT2, etc.
The pulse delay can be from about 2us to 20ms. The pulse width can be any
value from 1us to 20ms.
Syntax:
RTf,p,d,a
Where:
f = output channel (0 to 7)
p = pulse width (microseconds)
d = delay from trigger to pulse (microseconds)
a = output current (A)
Example of Pulsed Output Mode
Set channel 2 to pulsed output at 4A, pulse a continuous current of 0.5A
on output channel 3 for 1ms after a delay of 500us
RT2,1000,500,4