wrap Teflon
thread tape around the NPT threads
three times and install it into the DUT.
Install the Device Under Test (DUT). The DUT will
be connected to the test line with the remaining
flexible 5 in. nipple. This nipple can be installed either
in the 3-way tee, or in the port on top of the MPG2-H.
The tube can be slightly bent if necessary. Install the
nipple in the desired location. Tighten the gland in the
tee or MPG2-H, but do not tighten the gland that
connects to the DUT.
Plug the open test port: There is an open test port
either on the top of the MPG2-H, or on the 3-way
tee. Insert the stainless steel plug into a gland.
Insert the plug and gland into the open port. Do not
tighten.
Prime the system, tighten fittings: Pump fluid
through the nipples by using the MPG2-H. See
the appropriate section of the MPG2-H Operation
and Maintenance Manual for this procedure.
Make sure that all air is purged from the test lines.
There are 3 loose fittings: one on the DUT, one on
the RPM4 or PG7302, and the other on the plug.
Tighten each of the loose fittings when oil without
bubbles begins to flow through it.
t 3367676
Documen
DH Instruments,
a Fluke Company
4765 East Beautiful Lane
Phoenix AZ 85044-5318
USA
Figure 1: Nipples installed in 3-way tee, elbow and union
W Warning
• The maximum operating pressure of
the variable volume is
(15,000 psi) in MPG2-H-HV and 200
MPa (30,000 psi) in MPG2-H. Do not
generate pressure greater than the
maximum operating pressure.
• High pressure fluids are potentially
hazardous.
Energy
pressurized fluids can be released
unexpectedly and with extreme force.
High pressure systems should be
assembled and operated only by
personnel who have been instructed
in proper safety practices.
• ALWAYS use external tubing and
fittings rated for pressures equal to
or
greater
than
pressure which MPG2-H will be used
to generate.
100 MPa
stored
in
the
maximum
081001
Tel 602.431.9100