Olympus B2-BDS-2 Sorun Giderme ve Ayarlama Kılavuzu - Sayfa 23

Mikroskop Olympus B2-BDS-2 için çevrimiçi göz atın veya pdf Sorun Giderme ve Ayarlama Kılavuzu indirin. Olympus B2-BDS-2 46 sayfaları. For electrical base

Olympus B2-BDS-2 Sorun Giderme ve Ayarlama Kılavuzu
B.
TROUBLESHOOTING (B2-BDS-2)
No. B-6
4.
Power
Supply
4-1
Power
is not supplied
even
after the main switch is turned on.
SWI03
TIOt
LAMP
UYPC65
~-----J_2_o_tJ:~~~=:=:=:=:=:=:=:=:=:=:=:~====================~
Jt02
Causes
1.
The breaker functions.
2.
Blown
main fuse FlOl
3.
Defective UYPC65
4.
Defective UYPC64
5.
Defective transformer Tl02
6.
Defective transformer TlOl
7.
Defective main switch SW102
8.
Defective primary voltage selector switch SW103
9.
Defective breaker SWlOl
10.
Defective noise filter RClOl
11.
Defective spark gap RC102/RC103
12.
Defective
AC inlet
JlOl
13.
Defective AC cord (UYCPlS in lOOV channel)
Is the breaker functioning?
NO
YES
0-----c~~~-®
Press the breaker button.
Is power
supplied
normally?