Olympus B2-BDS-2 Instrukcja rozwiązywania problemów i regulacji - Strona 23
Przeglądaj online lub pobierz pdf Instrukcja rozwiązywania problemów i regulacji dla Mikroskop Olympus B2-BDS-2. Olympus B2-BDS-2 46 stron. For electrical base
B.
TROUBLESHOOTING (B2-BDS-2)
No. B-6
4.
Power
Supply
4-1
Power
is not supplied
even
after the main switch is turned on.
SWI03
TIOt
LAMP
UYPC65
~-----J_2_o_tJ:~~~=:=:=:=:=:=:=:=:=:=:=:~====================~
Jt02
Causes
1.
The breaker functions.
2.
Blown
main fuse FlOl
3.
Defective UYPC65
4.
Defective UYPC64
5.
Defective transformer Tl02
6.
Defective transformer TlOl
7.
Defective main switch SW102
8.
Defective primary voltage selector switch SW103
9.
Defective breaker SWlOl
10.
Defective noise filter RClOl
11.
Defective spark gap RC102/RC103
12.
Defective
AC inlet
JlOl
13.
Defective AC cord (UYCPlS in lOOV channel)
Is the breaker functioning?
NO
YES
0-----c~~~-®
Press the breaker button.
Is power
supplied
normally?