Olympus bx40 Talimatlar Kılavuzu - Sayfa 18

Mikroskop Olympus bx40 için çevrimiçi göz atın veya pdf Talimatlar Kılavuzu indirin. Olympus bx40 33 sayfaları. Replacement illuminator
Ayrıca Olympus bx40 için: Kurulum Talimatları (4 sayfalar)

Olympus bx40 Talimatlar Kılavuzu
Fig. 23
Fig. 24
Fig. 25
1. Slightly loosen the Stage clamping screw 0.
2. The Stage tan be rotated both to the left and to the right by turning it
with the Stage clamping screw.
Sr A click may be heard and felt during rotation.
However,
this is
due to the construction
of the substage and does not indicate a
malfunction.
0 The angle of rotation varies depending on the Position of the Stage knobs.
Specifications are identical for both the Standard type and long
type knobs.
0 By lowering
the Position
of the substage,
the microscope
will
accommodate specimens with maximum heights of 40 mm. This is useful
when observing metallurgical specimens and other thick objects.
1. Lower the Stage to the lower limit, then remove the Stage from the
microscope. (See page 3)
2. Using the Allen screwdriver, loosen the substage bracket clamping screw
0 and remove the substage. (Fig. 24)
3. Turn the coarse adjustment knob and raise the focusing block @ to where
the stopper screw @ in the arm becomes visible. (Fig. 25)
4. Using the Allen screwdriver, loosen and remove the upper stopper screw
0.
5. Reattach substage bracket and Stage.
0 Store the removed stopper screw @ in a safe place so that it will not
be lost, if needed again.
13
0