Olympus BX61 Talimatlar Kılavuzu - Sayfa 25

Mikroskop Olympus BX61 için çevrimiçi göz atın veya pdf Talimatlar Kılavuzu indirin. Olympus BX61 40 sayfaları. Motorized system microscopes

Olympus BX61 Talimatlar Kılavuzu

3-6 Immersion Objectives

# Be sure to use the provided Olympus Immersion oil.
Fig. 36

3-7 Objectives with Correction Collar

@
Fig. 37
1
Using Immersion Objectives
1. Focus on the specimen using all objectives, starting from the lowest-
power objective to higher-power objectives.
2. Press the stage escape button to lower the stage, then place a drop of
provided immersion oil onto the specimen at the area to be observed.
3. Turn the revolving nosepiece to engage the immersion objective.
4. Press the stage escape button to raise the stage, then turn the focus
adjustment knob to focus on the specimen.
# Since air bubbles in the oil will affect the image quality, make sure
that the oil is free of bubbles.
a. To check for bubbles, remove the eyepiece and fully open the field and
aperture iris diaphragms, then look at the exit pupil of the objective inside
the observation tube. (The pupil should appear round and bright.)
b. To remove bubbles, turn the revolving nosepiece toward the shorter ob-
jectives and repeatedly defocus and refocus the oil immersion objective.
}If the condenser engraving shows a numerical aperture (NA) of 1.0 or
higher, the number applies only when oil is applied between the slide
glass and the top surface of the condenser. When oil is not present, the
NA is about 0.9.
5. After use, remove oil from the objective front lens by wiping with gauze
slightly moistened with absolute alcohol.
!Caution in use of immersion oil
If immersion oil enters your eyes or contacts with your skin, immediately
take the following treatment.
Eyes: Rinse with fresh water (for 15 minutes or more).
Skin: Rise with water and soap.
If the appearance of the eyes or skin is altered or pain persists, immedi-
ately see your doctor.
}If the cover glass thickness is not 0.17 mm, the objectives cannot mani-
fest their performances. If a correction collar equipped objective is used
in this case, the difference in thickness can be compensated for by ad-
justing the collar.
Adjustment Procedure
· If the cover glass thickness is known, set the correction collar @ to that
value. (Fig. 37)
· If the cover glass thickness is unknown, adjust the correction collar @
and focus adjustment knob alternately until the positioning with the highest
resolution is obtained.
# Be careful not to touch the correction collar @ when turning the
revolving nosepiece.
BX61
(Fig. 36)
21