Safety Symbols
The following symbols are found on the microscope. Study the meaning of the symbols and always use the equip-
ment in the safest possible manner.
Symbol
Indicates that the surface becomes hot, and should not be touched with bare hands.
Before use, carefully read the instruction manual. Improper use could result in personal injury to
the user and/or damage to the equipment.
l
Indicates that the main switch is ON.
Indicates that the main switch is OFF.
Warnings
Warning engraving/stickers are placed at parts where special precaution is required when handling and using the
microscope. Always heed the warnings.
Warning engraving
position
9. When using a photographic unit or TV camera, be sure to observe the
total weight limit of the modules added to the standard combination as
shown below.
Illuminator
BXFM-ILHS
Mounting Holder
Standard
Combination
U-KMAS, U-TR30-2,
Weight (Module
U-LH100-3, revolving
Names)
nosepiece, objec-
tives, eyepiece
Additional Load
Limit
· With the combination using the BXFM-ILH illuminator mounting holder,
do not use the U-ST stand because this makes the system unstable.
· The system can be stabilized by using the SZ-STL large stand, but the
image may fluctuate or blur irregularly when a 50X or higher-power ob-
jective is used.
· It is recommended to fabricate an exclusive clamping tool which uses
the support mounting hole (P. 20) or module mounting screw hole of the
BXFM-F focusing unit.
Explanation
Lamp housing
(Warning against high temperature)
BXFM-ILH
5.2 kg
7.5 kg
BX-RLA2, U-TR30-2,
U-LH100-3, revolving
nosepiece, objec-
tives, eyepiece
4.0 kg
When the BXFM-
2.8 kg
ILHSPU counter
spring is used
BXFM
2