Olympus PSD-10 Talimatlar Kılavuzu - Sayfa 26
Tıbbi Ekipman Olympus PSD-10 için çevrimiçi göz atın veya pdf Talimatlar Kılavuzu indirin. Olympus PSD-10 28 sayfaları. Electrosurgical unit
9
M A I N T E N A N C E
9-1 Replacement of Fuse
Turn off power switch and unplug
from hospital grade receptacle.
Fuse holder
9-2 Cleaning of Ventilation Grills
9-3 Periodic Inspection
If the PSD-10 ceases to operate and the malfunction cannot be isolated, check
the fuse holder on the rear of the PSD-10 after unplugging the power cord from
the hospital grade receptacle.
Be sure to unplug the power cord from the AC mains.
If the fuse is found to have blown, replace it with a new fuse making sure
that the requirement for supply voltage fluctuation in 1-2 Main Specifications,
is met.
In case the spare fuse has been expended, use a commercially available
equivalent (rated value is indicated near the fuse holder).
After replacement, if the new fuse blows again, contact an Olympus service
center.
If dust and dirt have collected on the ventilation grills, clean using a vacuum
cleaner.
Periodically inspect the PSD-10 following the procedure in Section 6 IN-
SPECTION.
lf any irregularity or abnormality is suspected, contact your Olympus repre-
sentative.
Repair must be made by an authorized Olympus service center, and should
not be attempted by non-Olympus service personnel.
Non-Controlled Document