Olympus CKX41 Onarım Kılavuzu - Sayfa 33
Laboratuvar Ekipmanları Olympus CKX41 için çevrimiçi göz atın veya pdf Onarım Kılavuzu indirin. Olympus CKX41 43 sayfaları. Reflected fluorescence system
Ayrıca Olympus CKX41 için: Genel Bakış (7 sayfalar), Broşür ve Teknik Özellikler (4 sayfalar), Talimatlar Kılavuzu (36 sayfalar)
CKX31/CKX41
2- 3 Parfocality adjustment for CKX31
Relay lens ass'y consists of three parts, lens frame-1(A), frame-2(B) and lens frame-3(C).
(For the composition of lenes, refer to the Fig.1 and Fig.2.)
Set the jigs to the microscope and check the parfocality. (Refer to D-5.)
Adjustment procedure:
(1) Remove the stage part, and take off the covers. (refer to the disassembly on C-1.)
(2) Remove the screws(*1) and remove the part of relay lens ass'y and BI tube.
Screw: AB3X14SA(*1), Washer: BNW3SA 4pcs. each
(3) Loosen the screw(*2) and turn the relay lens ass'y counterclockwise to remove it.
Screw: ACU3X6SA(*2) 1pc.
(4) Remove the lens frame-1 ass'y(A) and insert appropriate washer(selection).
(5) Assemble the parts in reverse order of disassembly.
(6) Set the jigs, focus on the reticle of KN0048 and the specimen in KN0041.
At this time, read the difference from the index on the helicoid scale.
Relay lens Ass'y
(B)
C)
Fig. 1
D. REPAIR PROCEDURE
BI tube
*2
(A)
(D)
Selection: washer
AE092300 ( t=1mm) Standard
AE092100 ( t=0.5mm)
AE092200 ( t=0.7mm)
AB129900 ( t=0.05mm)
AB130000 ( t=0.1mm)
AB130200 ( t=0.3mm)
Fig. 2
*1
For setting the jigs,
refer to D-5.
D-6
2
Scale
Standard:
+/- 0.3mm
(without 3 graduations)
*For adjustment amount,
refer to next page.
Washer insertion
screwed to BI tube