AccuMed ABP801B Kullanım Kılavuzu - Sayfa 15
Kan Basıncı Monitörü AccuMed ABP801B için çevrimiçi göz atın veya pdf Kullanım Kılavuzu indirin. AccuMed ABP801B 15 sayfaları. Wrist blood pressure monitor
applicable RF compliance level above, the MODEL ABP801 should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the
MODEL ABP801.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
portable and mobile RF communications equipment and the MODEL ABP801
The MODEL ABP801 is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the MODEL ABP801 can help
prevent electromagnetic interference by maintaining a minimum distance between portable
and mobile RF communications equipment (transmitters) and the MODEL ABP801 as
recommended below, according to the maximum output power of the communications
equipment.
Rated
maximum
output power
of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Recommended separation distances between
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
0.12
0.38
1.2
3.8
12
80 MHz to 800 MHz
0.12
0.38
1.2
3.8
12
13
800 MHz to 2.5 GHz
0.23
0.73
2.3
7.3
23