7 .  i
nstallation
a)  Install the level meter in a vertical position into the upper lid of the tank (vessel) or reservoir 
using a lug, a fastening nut or a flange in such a way that the ULM axis is perpendicular to the 
level of the measured liquid (Fig. 1).
b)    M inimum dimension parameters when installing into 
a lid or a ceiling of a tank are listed in Fig. 3.  
c)      W hen installing in an open channel (reservoir, drain 
and the like), install the level meter onto a console to 
the expected maximum level as close as possible.
d)      I n compliance with the measurement principle, no 
signals reflected in the area directly below the level 
meter (the so-called dead zone) can be evaluated. 
The  dead  zone  (Fig.  2)  determines  the  minimum 
distance possible between the level meter and the 
highest  surface  level.  Medium  minimum  distance 
parameters  are  listed  in  chapter  on  „Technical 
specifications".
e)      I t is necessary to install the level meter in such a way 
that  the  bin  level  does  not  interfere  with  the  dead 
zone  when filled up to the  maximum. If measured 
level interferes with the dead zone, the level meter 
will not work properly.
Fig. 2: Level meter dead zone
f)  In case the maximum surface level in the tank interferes with the dead zone, the level meter 
has to be mounted into a higher installation neck. Subsequently, the tank can be filled nearly up 
to the maximum volume. The neck's inner surface has to be even and smooth (without edges 
and welded joints), the inner edge should be rounded in the position point where the ultrasonic 
waves leave the pipe. Choose the largest possible neck's diameter, but keep the neck's height 
as low as possible. Recommended dimensions of the neck are listed in Fig. 4.
Fig. 4: Possible installation of the installation neck
© Dinel, s.r.o.
ULM–53L
instrUctions
a – Neck heigh
b – Neck width
m – Dead zone
Fig. 1: Recommended installation
ULM–53L–02 ; 10
ULM–53L–06
ULM–53L–20
Fig. 3: Installation distance from the tank wall
ULM–53L–02 ; 06
ULM–53L–10
ULM–53L–20
in the tank
d > 1/12 c
(Min. 200 mm)
d > 1/8 c
(Min. 200 mm)
d > 1/10 c
(Min. 200 mm)
d – Distance from tank wall
c – Measuring range
m – Dead zone
a < 3 b
b > 100 mm
a < 1,5 b
b > 100 mm
a < 1,5 b
b > 150 mm
7