9. Installing seat with the rear opening (accessory)
The seat comprises two halves that are snapped in place with the aid of clamps
on the seat's underside.
10. Sanitary pan
The sanitary pan is slid into position from behind between the seat and wire
guides as shown in fig. 9. Make sure the sanitary pan is slid into position above
the wire guides. Also make sure that the pan assumes the right position in relation
to the cutout/seat and that the pan's rear hook locks into the guide (see figs. 9
and 10).
11. Footrest / Heel restraint
The foot rest can be easily set to the preferred height, swung aside (lift/swing) or
removed (lift away). Set footrest height by moving the green clip to the preferred
hole.
12. Castors
The Easy Hygiene Chair can feature 2 different types of wheels. The images show
the options.
A) Single wheel 125mm
The Hygiene Chair has 4 wheels (full brakes). The wheel on the right-hand front
side (grey brake pad) also features a directional lock (see image 14 - brake
position 1 - directional lock, brake position 2 - full brake).
B) Double wheel, 100mm
The Hygiene Chair has 4 wheels, of which the one on the righthand front side
features a directional lock (see image 15) and the other 3 are fitted with full
brakes.
When the directional lock is activated, wheels cannot be turned. This is
particularly useful when moving heavier patients. The chair is thus significantly
easier for staff to push and control. The directional lock should be released
when manoeuvring in tight spaces. The other wheels (x 3)/green brake pad are
conventional wheels that lock both directionally and rotationally when the brakes
are applied (see image 16, image 17).
13. Calibrating the actuators
The actuators are normally calibrated automatically every time the chair reaches
its end position (min height, max tilt). IF the chair still does not run parallel, or if the
tilt function does not seem to work, an ACTIVE calibration should be performed.
Do this by holding down the <slow/up>+ <slow/down> buttons simultaneously.
After around 15 sec., the chair will move to its end position i.e. minimum height
and maximum rearwards tilt Hold down the buttons until the end positions are
reached. Release the buttons and wait five sec. The chair is now calibrated and
can be used again.
Calibration should NOT be done while user is in the chair. Note that all three actuators move during the calibration
phase. If not, stop calibration immediately and contact customer service/GATE (or your local dealer).