Select the icon for Advanced Settings and input the password 112112.
1. Blow: This icon under advanced settings allows the user to change
the blow time (in seconds), initiate flow rate (in L/M) and minimum
volume (in L). These are already set for optimal performance; it is not
recommended that they be changed. Default settings are 8L, 3 s and
1.2L.
2. Default: Select yes under this icon to return instrument to Default
settings. NOTE: Settings have been optimized and it is not
recommended to return to Default settings.
3. Unit: Set the measurement unit and BAC/BrAC ratio. If unit is
changed, the instrument does not need to be recalibrated.
4. Password: The default passwords for advanced settings and
calibration can be changed. Select "Cal" or "Advanced Setting" to
change the password. Enter the new password and select
confirm. Screen will display "Password Setting Succeed".
NOTE: Instrument requires a 6-digit password. If less than six
digits are used, password will not store properly.
Warning: If you change the password, write it down. The new
password must be sent in with your unit if it should need
service.
You can also choose whether to save Passive Screening Test Results by checking the box,
the default is Not Saved.
5. Delete: This option allows for the deletion of ALL saved test records
stored in the Jupiter. Once you select "Delete All Records", they
cannot be restored.
6. Field:
i. Nine different fields can be set.
Advanced Settings:
to