CUSTOM ULTRASONICS 20000 Operatör El Kitabı - Sayfa 2
Depolama CUSTOM ULTRASONICS 20000 için çevrimiçi göz atın veya pdf Operatör El Kitabı indirin. CUSTOM ULTRASONICS 20000 7 sayfaları. Endoscope storage cabinet
Ayrıca CUSTOM ULTRASONICS 20000 için: Operatör El Kitabı (11 sayfalar)
CUSTOM ULTRASONICS, INC.
M
ODEL
P
D
HYSICAL
ESCRIPTION
Model 20000 Endoscope Storage Cabinet
The
channels with forced air during storage. It is fabricated of painted steel to provide an environment
for endoscope storage that is free of water-absorbent materials [e.g.: laminated wood] to minimize
the risk of endoscope bacterial and/or fungal growth.
The cabinet has been designed to accommodate up to (8) flexible endoscopes. The actual number
that the cabinet accommodates will depend on the configuration or "design" of the endoscopes.
A gasket surrounds the door opening. When the door is closed (by the adjustable and lockable
handle), the gasket is compressed providing a sealed enclosure.
At the top of the cabinet is an enclosure containing an air pump for the circulation of air through
the channels of the endoscopes. Next to the air pump, on the other side of the upper enclosed area,
is a chamber for desiccant bags that are used to absorb moisture from the re-circulated air. Two re-
sealable plastic containers, each containing 6 desiccant bags, are provided with the initial Cabinet
purchase. One package of (3) Humidity Indicator Cards are also provided. (Additional bags and
cards are available – Desiccant bags Part # 20001 – Humidity Indicator Cards Part #20003)
A series of "vents", located on the bottom of the desiccant chamber, allows for air movement
through the desiccant and into the air pump for re-circulation through the endoscopes' internal
channels.
By opening the desiccant chamber door, the bags of desiccant can be removed, regenerated and
replaced. [Refer to the maintenance section later in this manual.]
Below the desiccant bags and air pump chamber, at the top of and inside the main chamber of the
cabinet, is a manifold containing several bulkhead fittings. Adapters are used to connect the
endoscope to these bulkhead fittings. Circulation of air through the endoscopes' internal channels
occurs via these adapters. Although a relief valve protects the air pump in the absence of adapters
plugged into the bulkhead fittings, one adapter must be connected to the manifold.
A holding compartment for the endoscopes' operating buttons is located inside the door of the
cabinet.
The storage cabinet's ON/OFF main switch and the adjustable programmable timer are located
outside of the cabinet on the right front wall.
Each
Model 20000 Endoscope Storage Cabinet
cabinet may reduce the incidence of endoscope theft.
SHEET
2 OF 7
20000 E
NDOSCOPE
:
REVISION
A
S
C
TORAGE
ABINET
continuously purges each of the endoscopes'
is equipped with a lock. Routinely locking the
M
20000 E
ODEL
NDOSCOPE
S
C
TORAGE
ABINET
O
'
M
PERATOR
S
ANUAL