User instructions and information
Connecting accessories
Connect the instrument, cables, electrodes and, if necessary, grounding according
to the measurement method to be used.
Measurement methods
Point-to-point measurement (Rp-p)
Connect the BNC plug ends of the wires to the
instrument and the black shielding wire connectors
to the black sockets of the electrodes and the red
measuring wire connectors to the red sockets of the
electrodes.
Measurement of resistance to ground (Rg)
Connect the BNC plug ends of the wires to the
device and the connector of one of the black
shielding wires to the black socket of one of
the electrodes. To the red socket of the same
electrode, connect the red wire connected to
the black shielding wire already in use. Connect
the other red measuring wire to the ground.
The use of the other black shielding wire is not
necessary for this measurement method.
Surface resistance measurement (Rs)
Connect the BNC plug ends of the wires to the device and the
two red measuring wires to a ring electrode. (Ring electrode is
not provided in the AIJGO-61 product package.) The use of the
two black shielding wires is not necessary for this measurement
method.
Measurement of volume resistance (Rv)
Connect the BNC plug ends of the wires to the device and the
connector of one of the red measuring wires to the red socket of
one of the electrodes. Connect the other red measuring wire to
the conductive material under the measured item. The use of the
black two shielding wires is not necessary for this measurement
method.
5