Batteries and charging system

Charging and handling the battery

The battery charger can be used as a table-top
unit or can be fixed to the wall. The Charger must
be placed or installed in a way that makes it easy
to disconnect mains cable plug. Take care not to
damage the cable.
Charger light
Insert the battery into the charger. The light on the
charger will come on.
LED Mode
No bat
6
6
/
6
<
/
Error!

Battery capacity

The LED on the hand control and on the battery
holder will illuminate when the battery power
falls below a certain level, indicating that charging
is needed. A sound signal will also be heard. If
the battery goes flat in the middle of a lifting
manoeuvre, there will always be sufficient power
to lower the patient down again.
BM08101 Molift MOVER 180 English - Rev C 19.12.2013
New batteries will not work at full capacity
until they have been fully charged and run
empty several times.

Electronics

The lift counts the number of lifting manoeuvres
performed. This information can be read using the
Molift Service Tool, which will indicate when the lift
needs a service, e.g. a new lift motor. Please also
refer to the Maintenance chapter.
The electronics have an energy-saving function
that shuts the system down after approx. ten
minutes of inactivity. All the lights will go off. The
electronics will be activated by pressing one of the
operating buttons.
If the lift is in constant use for a long period of time,
there is a risk of the electronics overheating and the
lift stopping. Allow the lift to cool down sufficiently
before using it again.
Discarded lifters and batteries should be
handled as electronic waste and collected
seperately according to local laws and
regulations.
Molift Mover 180 should not be run
constantly for more than 2 minutes (with
maximum load), and rest for minimum 18
minutes.
Battery light
Service light
Electrical
emergency
lowering
function
Emergency
stop
Page 9 of 16