Programming the AFI4 / AFI5 with the FlexProgram
Setting of Temperature compensation for each range.
This will be factory set to 2%/°C as default.
If the box Advanced is ticked, it will be possible to enter an
other target temperature. This is as default set to 25°C, which is the
most common target temperature for most media. If another target
temperature is wanted, it can be entered.
NOTE:
This is not the specific media temperature, but the
"0-point" for the temperature compensation.
When the "Advanced" is ticked the "Compensation temperature
source" is accessible.
You can enter a value to the column %/K
curve.
These fields are not necessary to change for standard measurement.
Be careful if changing the default values, as wrong values in those
fields may course wrong measurement.
It is possible to select from where the temperature measuring shall
come from
Sensor tip
HART
Fixed
Screen Input (when concentration is selected in DTM setup)
No settings in this screen will be available as standard.
When user defined concentration is selected in screen concentration,
it will be possible to adjust temperature concentration.
The default setting will be 2%/°C.
Advanced settings
Here only "Compensation temperature source" will be available.
Screen Concentration
There is 4 selectable medias, for which the calibration curves
is default entered from factory.
NaOH
(caustic soda)
NaOH
-
HNO
(Nitric acid)
3
HNO
3
For user defined, please see next page.
www.baumer.com
2
you will create a unlinear
(built in Pt100)
(external input via HART)
(no temperature measurement)
0...15% by weight
25...50% by weight
0...25% by weight
36...82% by weight
CombiLyz, AFI4 / AFI5