1) Before connecting the USB cable, always power up the camera using the PSU supplied.
2) The camera has internal hardware protection from overheating. Should the body of the camera reach
>40º C, a buzzer will sound and the power to the cooler will be disconnected until a safe temperature is
reached.
3) For certain cameras, using a PS-25 power supply in Deep Cooling mode is designed for water cooling
only. It may, however, be possible to use this PS-25 power supply with air cooling depending on the
ambient air temperature. Typically the ambient air temperature must remain below 20 ˚C to dissipate the
extra heat generated in this mode of operation.
4) When a safe temperature is reached, the camera will Auto-Reset. If the buzzer sounds, check for the
following:
 Any obstructions in the air or water flow(s)
 Ambient air temperature & water temperature
 The temperature set in the Temperature screen. Set a lower temperature if required
Should you have any problems or queries, please contact the nearest Technical Support Engineer as