7.8. Replacing the Control Board
The control board is accessible once the cabinet has been removed. Make certain
that all wire harnesses have been disconnected.
The control board has capacitors that will remain charged for a period after the
centrifuge is unplugged. Make certain to use standard precautions for handling
potentially charged capacitors when working with the control board.
The control board is held in place with six screws. Use a #2 Philips driver to remove
the screws. The board is now free to remove.
To install the control board, make certain that the six posts line up with the holes
in the control board and use a # 2 driver to install the screws.
7.9. Removing the motor
Remove the cabinet assembly as previously described.
Flip the base assembly up‐side down and remove the three screws that are
securing the exhaust cover and set aside.
Remove the six screws that hold the guard bowl
Lift the base assembly off the guard bowl and set it aside
Disconnect the motor wire connectors.
Remove the ground connection with a nut driver.
Flip the guard bowl.
The motor is held in place with four #8 Nylok nuts. Remove them with an 11/32"
nut driver.
Slide the motor out of the guard bowl's motor well.
DRUCKER DIAGNOSTICS
HORIZON 6 SERVICE MANUAL
SM024
REV A
7