LED indications on front of module
Red LED
Meaning
Internal fault: power supply not
sufficient, out of range offset or
Fast flashing
reference. Error on reading or
writing in flash (at the start or on
threshold setting).
Dip-switch setting error
Slow flashing
Set threshold in progress
Disconnected thermocouple, out
of range input or temperature
Steady light
compensation
Output limiting in progress
* In this modality the output signal represents the value of the threshold.
Yellow LED
Meaning
ON
The auxiliary output is closed
OFF
The auxiliary output is open
Electrical connections
The module is designed for spring cage clamp
electrical connections.
1. Strip cables by 0.8mm.
2. Insert screwdriver in the square hole and press
until the cable lock spring opens.
3. Insert cable in the round hole.
0.2 to 2.5mm
2
4. Remove screwdriver and ensure cable is tightly
8mm
fastened in the terminal.
Power supply
+
19.2 to 30VDC
-
There are three ways to power the DSCP6x series of signal converters.
1. Connect the 24VDC power supply directly to terminals 7 (+) and 8 (-)
of each module.
2. Connect power to one signal converter and use the expandable Power-
Bus connector to distribute power to a maximum of 16 adjacent modules.
The bus can be supplied from any of the modules, but the total current
consumption of the bus must be less than 400mA. Higher consumption
values can damage the module. An appropriately sized fuse must be
connected in series with the power supply.
3. Use the DSCP70 Power Supply Connection Module and the expandable
Power-Bus connector to distribute power to a maximum of 75 modules.
The DSCP70 is designed to protect the modules connected via bus against
overvoltage loads. The bus connector can be provided with power using the
DSCP70 module if the total consumption of the bus is less than 1.5A. Higher
consumption values can damage both the module and the bus. An appropri-
ately sized fuse must be connected in series with the power supply.
Dataforth Corporation
3331 E. Hemisphere Loop • Tucson, AZ 85706 USA • Toll Free: 800-444-7644 • Tel: 520-741-1404 • Fax: 520-741-0762 • Email: [email protected] • www.dataforth.com
SD1145 02/11 ©2011 Dataforth Corporation, All Rights Reserved ISO9001:2008 Registered QMS
Output fault
Yes
Yes
No *
Yes
No
Red LED
Set threshold
Yellow LED
5
1
6
2
3
7
8
4
Power
supply
Input
The module accepts input from the following types of thermocouples:
J, K, R, S, T, B, E, N.
The use of shielded cables is recommended for the electronic connections
Output: Voltage / Current connections
The use of shielded cables is recommended for the electronic connections.
+
V / I
-
Auxiliary output
The auxiliary output is designed to drive an indicator, a relay of greater
power, or the input of a supervisory control system.
+
V / I
5
1
-
Output
6
2
Input
3
7
8
4
Power
supply
5
1
Output
6
2
+
Input
7
3
TC
8
4
Power
supply
5
5
1
Output
6
6
2
Input
3
7
7
8
8
4
Power
Power
supply
supply
AUX
5
1
Output
AUX
6
2
Input
3
7
8
4
Power
supply
AUX
1
Output
AUX
2
Input
3
4
3