dymax ACCU-CAL 50-LD Посібник користувача - Сторінка 17
Переглянути онлайн або завантажити pdf Посібник користувача для Вимірювальні прилади dymax ACCU-CAL 50-LD. dymax ACCU-CAL 50-LD 20 сторінок. Led radiometer
Maintenance
The ACCU-CAL™ 50-LED was designed to operate with minimum maintenance. Follow the schedule below to
assure top unit performance.
Calibrate the instrument at least annually. Calibration service is available through Dymax Customer Service
or Dymax Product Repair.
Change the batteries when a low battery warning is received. The ACCU-CAL™ 50-LED uses two AA-type
batteries. The battery compartment is on the back of the instrument.
Figure 19. Battery Compartment (Closed & Open)
Keep the detector head's sensing element clean and free of contaminants. The detector head may be
cleaned with a clean tissue wetted with isopropyl alcohol.
Spare Parts and Accessories
Spare Parts
Item
BlueWave® QX4™ Optic Adapter Upgrade Kit (includes optometer calibration update)
Adapter Kit – Flood to Spot Model (includes parts listed below)
Line Array Adapter Upgrade Kit (Includes optometer firmware, calibration update,
and Line Array Adapter)
Lightguide Simulator
8-mm Lightguide Adapter
5-mm Lightguide Adapter
3-mm Lightguide Adapter
Line Array Adapter
2 mm Hex Wrench
2 mm Screws
Dymax ACCU-CAL™ 50-LED Radiometer User Guide
17
Part Number
42218
39554
43383
38408
39558
39557
39556
43364
38656
38881