Maintenance
The ACCU-CAL™ 50-LED was designed to operate with minimum maintenance. Follow the schedule below to
assure top unit performance.
Calibrate the instrument at least annually. Calibration service is available through Dymax Customer Service
or Dymax Product Repair.
Change the batteries when a low battery warning is received. The ACCU-CAL™ 50-LED uses two AA-type
batteries. The battery compartment is on the back of the instrument.
Figure 19. Battery Compartment (Closed & Open)
Keep the detector head's sensing element clean and free of contaminants. The detector head may be
cleaned with a clean tissue wetted with isopropyl alcohol.