dymax ACCU-CAL 50-LD Panduan Pengguna - Halaman 16

Jelajahi secara online atau unduh pdf Panduan Pengguna untuk Alat Ukur dymax ACCU-CAL 50-LD. dymax ACCU-CAL 50-LD 20 halaman. Led radiometer

16
Dymax ACCU-CAL™ 50-LED Radiometer User Guide
6. Press and release the Light Source button to select the light source being measured. For the BlueWave®
QX4™, use the "spot" setting. Repeatedly press the Light Source button until the correct focus lens size
appears on the top line. The options are:
3-mm Spot - For use when measuring a BlueWave® QX4™ Wand with the 3-mm Focus Lens installed
5-mm Spot - For use when measuring a BlueWave® QX4™ Wand with the 5-mm Focus Lens installed
8-mm Spot - For use when measuring a BlueWave® QX4™ Wand with the 8-mm Focus Lens installed
Figure 18. Spot Mode Screens (Shown in Peak Intensity Mode)
7. Press and release the Mode button to select the operating mode. The different operating modes are:
Peak Intensity - Shows the highest intensity in mW/cm
measurement and stores that value in the display.
Intensity - Shows the intensity in mW/cm
readout of intensity.
Dose - Shows the total energy dose in mJ/cm
Note: This measurement is time dependent.
8. Turn on the BlueWave® QX4™ to begin testing.
2
seen by the detector during the
2
at the detector during the measurement. This is a dynamic
2
at the detector during the measurement.