8
Returning Flexion & Extension
Adjustments to Factory Settings
A .
Unscrew and remove the brass Extension Screw (3mm Hex)
from the center of the Flexion Knob .
B .
Using the included Swing Adjustment Wrench (or a 9mm
wrench), gently screw the Flexion Knob clockwise until
it stops . Do not tighten .
C .
Now turn the Knob counter-clockwise 2 full rotations .
D .
While holding the Knob stationary,
screw the Extension Screw back into the Knob (fig. 1)
until one thread shows above the Knob (fig. 2) .
The knee's Flexion & Extension Controls
are now at their factory settings.
Fig. 1
Extension Screw
(3mm hex key)
(Posterior View)
Fig. 2
Extension Screw
(3mm hex)
Flexion Knob
(9mm)
DAW
®
DAW
Industries
© Copyright 2021, DAW Industries, San Diego, CA - All Rights Reserved 01082021
Flexion Knob
(9mm)
Swing Adjustment
Wrench
1 Thread showing
above Knob
For Technical Support call (800)242-8669
To Adjust Extension Assist:
A .
Loosen the Pylon Clamp Screw (5mm hex)
and remove the Pylon
Loosen the Lock Nut (12mm wrench)
B .
C .
Turn the Extension Assist Control Screw (5mm hex):
Clockwise to increase extension assist
Counter-clockwise to decrease extension assist
D .
Re-tighten the Lock Nut, Hand Tighten (12mm wrench)
After returning the pylon, re-tighten Pylon Clamp Screw to
11ft-lb (15Nm).
Pylon Clamp Screw
(5mm hex)
11ft-lb(15nm)
Lock Nut
(12mm wrench)
Hand Tighten
DAW
®
DAW
Industries
© Copyright 2021, DAW Industries, San Diego, CA - All Rights Reserved 01082021
Adjusting Extension Assist
Extension Assist Control Screw
(5mm hex)
For Technical Support call (800)242-8669