DayTronic IL70A Посібник - Сторінка 7
Переглянути онлайн або завантажити pdf Посібник для Блок управління DayTronic IL70A. DayTronic IL70A 15 сторінок. In-line dc strain gage conditioner
2
T
RANSDUCER
The Model IL70A I/O CONNECTIONS are via non-removable screw terminals which will accept wire sizes from
AWG 16 to 30. NOTE: The recommended transducer cabling would be eight wire, individually shielded,
twisted pair - wired as indicated (Fig. 1). Sense lines must be connected at the transducer (as recommended)
or at the IL70A screw terminals - as a minimum. Table 2 denotes screw terminal assignments
I/O Connector
Terminal
J1
J2
Fig. 1 Model IL70A Transducer Cabling
DC Strain Gage Conditoner Module
C
ONNECTIONS
Table 2 Model IL70A Pin Assignments
Screw
Analog Output
1
SHP
2
SHN
3
NC
4
Analog Common
5
Power Common
6
+ 24 Vdc
7
Shield
8
+ Excitation
1
+ Sense
2
- Excitation
3
- Sense
4
+ Signal
5
- Signal
6
Cal Sense
7
Shield
8
Terminal
Label
7 Wire confirguration (
4 twisted pair, shielded wire used.
Paired as follows:
4 Wire confirguration
2 twisted pair, shielded wire used.
Paired as follows:
Sense lines must be connected at the
instrument
TRANSDUCER CONNECTIONS
IL70A
Conditioner
Line
Function
+ Signal Output
Shunt Positive switch input
Shunt Negative swith input
No connection
- Signal Output
Power Supply Common
11 - 28 Vdc Input Power
Earth ground
+ Sensor Power
+ Sensor Power Sense
- Sensor Power
- Sensor Power Sense
+ Signal from Sensor
- Signal from Sensor
Calibration Shunt Sense
Earth ground
recommended
+ & - Excitation
+ & - Sense
+ & - Signal
Cal Sense and unused wire
Shield at the instrument only
+ & - Excitation
+ & - Signal
Shield at the instument only
)
IL70A.2