3. CALIBRATION
Calibration of the IL70A is accomplished by the conventional shunt technique, using an internally
installed calibration resistor, or via the "dead-weight" method.
Calibration Resistor. If a fixed resistor is shunted across one arm of a strain gage bridge, it produces an
unbalance equivalent to that of a particular value of mechanical input. If this Equivalent Input value is
accurately known, it can be used as a reference point for shunt calibration of the system. Upon
completion of installation of the transducer and its associated cabling, the user can:
Perform an overall dead weight calibration using a precisely known value
(1)
of mechanical input. The calibration can then be transferred to the installed
calibration resistor for convenience in subsequent checking.
Use an external shunt resistor, please refer to the notes on page one of this manual. The
(2)
installed calibration resistor is typically supplied by the transducer manufacturer or is
referenced on the calbiraiton sheet that accompanies the tranducer. Using a known shunt
resistorance can achieve a precisely known Equivalent Input allowing the instrument
sensitivity to be adjusted correctly. Note, this shunt resistor may reside in the transducer
itself and is accessed via the input pins of the sensor.
Determine the Equivalent Input value for the internally installed 59K calibration resistor,
(3)
knowing the transducer sensitivity, and adjust the instrument sensitivity accordingly. Place the
jumper on the IL70A to "INT" to enable the 59K Ohm resistor.
A one percent, 59-kilohm calibration resistor is installed in the IL70A at the factory. The installed
resistor can usually be used even though the transducer calibration data mentions some
other resistance value. In Section 4 of this manual, the techniques described above are
demonstrated. If, however, the installed value of calibration resistor is not appropriate
for the transducer and the measurement range to be used, an external "EXT" shunt resistor
should be installed. The external calibration resistor is installed from the "CAL Sense"
terminal to either the + Signal or - Signal teminals depending on your polarity direction
when using the provided "shunt" switch. Note: A variety of DC Load Cells and Slip Ring
Torque Transducers are supplied with the appropriate calibration resistor integral to the
transducer. When this type of transducer is used it will be necessary to place the jumper in
"EXT" and connect the "Cal Sense" line to the appropiate pin on the transducer.
This section contains the instructions for calibrating the IL70A. Included is a functional
description of the instrument front-panel (see Figure 2). To perform calibration, proceed
as follows with the sensor or calibration standard connected.
Turn power ON to the IL70A DC Power input terminal (11 to 28 Vdc input) The front-
(a)
panel indicator should light green to indicate the application of DC power. Allow 10
minutes of warmup for stabilization of transducer characteristics.
DC Strain Gage Conditoner Module
IL70A
Calibration
IL70A.3