Dickson KT6P5 Початок роботи - Сторінка 7

Переглянути онлайн або завантажити pdf Початок роботи для Диктофон Dickson KT6P5. Dickson KT6P5 7 сторінок. Remote sensing temperature recorder
Також для Dickson KT6P5: Посібники з швидкого старту (9 сторінок)

Dickson ensures that the Replaceable Sensors / Instruments listed below were developed, tested
and validated together.
Operation: The Instruments and Replaceable Sensor models listed above were specifically
designed to work together. No other sensors will work with the Instrument models listed. No
other Instrument will work with the Replaceable Sensors listed. The Replaceable Sensor must be
connected to the Instrument for the Instrument to operate properly.
Calibration: Only the Replaceable Sensor is calibrated. The sensor, and all calibration defaults
and adjustments are stored on the Replaceable Sensor. Accuracy is strictly controlled by the
sensor – no adjustments are made by the Instrument. Readings are sent from the Replaceable
Sensor to the Instrument for storage purposes only. Certificates of Calibration / NISTs are only
supplied for the Replaceable Sensor for this reason.
Recalibration: When a Replaceable Sensor is due for recalibration the existing Replaceable Sensor
can be replaced with a new Replaceable Sensor thus eliminating the need to return the entire
instrument for recalibration. Replaceable Sensors can be returned for recalibration as well.
Dickson KT6P Manual V3
1.800.561.8187
Replaceable Sensor Certificate of Validation
Replaceable Sensor Model: R400
Instrument Model: KT6P1, KT6P2, KT6P5
www.
7
.com