GUIDANCE AND MANUFACTURER'S DECLARATION
Electromagnetic Immunity
The Spinal Modulation Neurostimulation System is intended for use in the electromagnetic
environment specified below. The customer or the user of the Spinal Modulation Neurostimulation
System should assure that it is used in such an environment.
Immunity
IEC 60601 Test
Compliance
Test
Level
Level
Conducted RF
3 Vrms
3 V
IEC 61000-4-6
150 kHz to 80
MHz
Radiated RF
3 V/m
IEC 61000-4-3
3 V/m
80 MHz to 2.5
GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Electromagnetic Environment Guidance
Portable and mobile RF communications
equipment should be used no closer to any part
of Spinal Modulation Neurostimulation System,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
The recommended separation distance is a
minimum of 0.2 meter for transmitters of 80 MHz
to 2.5 GHz
Interference may occur in the vicinity of
equipment marked with the following symbol:
23
Recommended separation distances between portable and mobile RF communications
equipment and the Spinal Modulation Neurostimulation System
The Spinal Modulation Neurostimulation System is intended for use in an electromagnetic environment
in which radiated RF disturbances are controlled. The customer or the user of the Spinal Modulation
Neurostimulation System can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the System.
Separation distance according to frequency of transmitter
Rated maximum output
power of transmitter
(W)
150 kHz to 80 MHz
0.01
0.12m
0.1
0.37m
1
1.17m
10
3.70m
100
11.70m
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
(m)
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.12m
0.23m
0.37m
0.74m
1.17m
2.33m
3.70m
7.37m
11.70m