Axium Clinical Programmer Посібник користувача - Сторінка 16
Переглянути онлайн або завантажити pdf Посібник користувача для Медичне обладнання Axium Clinical Programmer. Axium Clinical Programmer 16 сторінок.
Pop-up Messages (Cont.)
Condition
Please specify a Stimulus
Attempt to program a set of
Response before programming
pulse parameters without a
Stimulus Response Name.
Programmer battery is low.
PDA battery reaches 30%.
Please recharge.
Programmer is booting after
Hardware reset Button
a reset. When you click OK
on the PDA was pressed,
the Programmer will switch
Programmer detected an
off. Press the Power button to
error or the Programmer is
restart.
launched for the first time.
Programmer storage space is
PDA storage is at full
low. Please contact your Spinal
capacity (log files are stored
Modulation representative for
on the PDA with each
maintenance.
significant operation such
as programming). Message
is displayed and file logging
continues by deleting old
log files to make room
for new log files. Normal
operation continues after user
acknowledgement.
Stimulation for one or more
One or more leads turned off
leads has been turned OFF.
due to Low Impedance.
Stimulation has been turned
Stimulation can be turned
OFF due to a magnet.
off by applying a magnet for
the duration specified by the
Magnet Stim On/Off Time
programmable parameter.
Stimulation has been turned
Stimulation can be turned off,
OFF.
either by the "All Stim OFF"
software button on the INS
programmer, or the Off switch
on the TNS.
Stimulator battery has reached
Battery has reached EOS
End of Service (EOS).
voltage. Stimulation is
Stimulation has been turned
disabled in order to preserve
OFF permanently.
power for RF communication.
Buttons
Resolution
"OK"
Select stimulus
response.
"OK"
Recharge as soon as
possible.
Press Power button.
"OK"
Contact your
Spinal Modulation
Representative.
"OK"
Perform impedance
measurement and re-
enable if within range.
Otherwise, investigate
lead integrity.
"OK"
Stimulation can be
turned back on by
applying magnet for
the same "Magnet
Stim On/Off Time".
"OK"
"OK"
Schedule replacement
of the Stimulator.
27
Pop-up Messages (Cont.)
Condition
Stimulator battery has reached
Battery has reached ERI
the Elective Replacement
voltage.
Indicator (ERI).
Stimulator clock is HH:MM. Do
The NS internal time is out
you want to update Stimulator
of sync with the Programmer
clock?
time.
The stimulator has been set to
The Programmer has
default values.
encountered an NS with
unreadable or invalid data.
Parameters have been set and
programmed to default values.
The stimulator has unreadable
NS has unreadable data (such
data. Please reconnect with
as Trim data)
programmer in FCE mode or
contact your Spinal Modulation
representative
The stimulator is in Upgrade
NS in Boot mode. All
Mode. Please reconnect with
stimulation is disabled.
programmer in FCE mode or
contact your Spinal Modulation
representative.
Unable to connect to
Can't connect to the NS.
stimulator. Make sure the
programmer is close to the
stimulator and try again.
Error messages may contain additional troubleshooting information such as " Error Code: 04-123,
BadParameterValue ". This is an aid for Spinal Modulation engineers to debug errors.
Buttons
Resolution
"OK"
Schedule replacement
of the Stimulator.
"Yes"
"No"
"OK"
Setup device
parameters as
desired.
"OK"
Contact your
Spinal Modulation
Representative.
"OK"
Contact your
Spinal Modulation
Representative.
"OK"
Move the programmer
above the Stimulator
in circular motions.
28