extinguishing  the  6V  LED.    If  the  setting  of  the  intensity  control  is  decreased  to  the  point  where  the  clamped  control 
signal  exceeds  the  threshold  waveform,  the  6V  LED  will  completely  extinguish.    The  operation  of  the  remaining  three 
comparators is similar to this one, differing only in their threshold voltages. 
Power LED 
Resistor R203 and LED "A" function as the power‐on indicator for the microscope.   
Errors in the Olympus Documentation 
In  the  course  of  researching  this  document,  two  errors  were  identified  in  the  schematic  diagram  published  in  the 
Olympus Research Microscope Series BH2 (BHS) Repair Manual.  Both of these errors are in the LED bar‐graph display 
circuitry. 
Bar‐Graph LEDs 
The  first  error  is  in  the  wiring  of  the  four  bar‐graph  LEDs,  as  shown  in  Detail  1  of  Appendix  2  and  Appendix  3  of  this 
document.  The Olympus manual shows these LEDs wired with their cathodes connecting to their respective comparator 
outputs  and  with  their  anodes  wired  to  their  respective  current‐limiting  resistors.    The  actual  circuit  configuration  (as 
observed on the main boards of functional BHTU equipment) has the anodes of the four LEDs connected to the filtered 
DC  power  supply  line  and  the  cathodes  connected  to  their  respective  current‐limiting  resistors.    This  is  a  minor 
discrepancy  and  either  configuration  would  work  equally  well,  but  the  actual  as‐built  configuration  is  shown  in  this 
document. 
Control‐Signal Clamping Network 
The  second  error  is  in  the  wiring  of  the  control‐signal  clamping  network,  as  shown  in  Detail  2  of  Appendix  2  and 
Appendix 3 of this document.  The Olympus manual shows this network fed from the full‐wave‐rectified output of the 
bridge rectifier, but in the actual circuit configuration (as observed on the main boards of functional BHTU equipment), 
this network is fed from the filtered DC power supply line.  This error is of greater consequence than the first, as the 
circuit published in the Olympus repair manual would not guarantee that the LEDs remain off around the zero crossings 
of the AC line. 
Removing the Electrical Base from the Stand 
To gain access to the electronics in the base of the BHT/BHTU stand, the electrical base must be removed from the base 
of the stand.  Be sure to take sufficient notes and plenty of photographs both before and during the disassembly process 
to make sure that  everything can be correctly reassembled later on.  Before removing  the electrical base, look at the 
right‐hand side of the unit, just above the intensity slider, and note whether or not a light preset control and switch are 
present.  The earlier units did not include these components, but the later ones did.   
Next, remove all the major components from the microscope stand (i.e., AC power cord, condenser, stage, eyepieces, 
viewing head, and objectives).  Doing so will make the stand easier to maneuver and will prevent inadvertent damage to 
these  components  during  disassembly  and  reassembly.    Be  sure  to  protect  these  components  from  dust  and  damage 
while they are not installed on the stand. 
After  removing  these  components,  cover  the  top  of  the  arm  (i.e.,  the  nosepiece  turret  and  the  exposed  mounting 
dovetail  for  the  viewing  head)  with  a  clean  plastic  bag  and  secure  this  with  a  rubber  band  to  keep  dust  out  of  these 
openings.    This  is  especially  important  on  BHTU  stands  which  have  an  optical  correction  lens  located  just  below  the 
viewing head to correct for the difference in tube length introduced with the reversed nosepiece. 
With the stand in its normal upright position, remove the lamp house from the back of the base.  The lamp house simply 
plugs into the base, and can be removed by grasping it and pulling it straight back.  Set the lamp house aside to prevent 
damage.    Do  not  touch  the  halogen  lamp  with  your  fingers,  as oils  from  your  skin  will  cause  premature  failure  of the 
bulb.  If the bulb is accidentally touched, clean it with isopropyl alcohol. 
Carefully  turn  the  stand  upside  down,  exercising  caution  to  prevent  impacts  to  the  focus  knobs  and  to  the 
condenser/stage holder.  Using a 5/32" Allen tool, remove the four hex socket‐head cap screws from the bottom of the 