TROUBLESHOOTING GUIDE

Under certain conditions, performance of the unit may be adversely affected by factors other than defects. If problems occur,
please review the following list and take remedial action as needed. If you cannot solve the problem after checking the entire
list, please contact your local Olympus representative for assistance.
Problem
1. Observation/illumination System
a) Bulb does not light.
b) Bulb is not lit.
c) Bulb operates, but field of view
remains dark.
d) Field of view is obscured or not
evenly illuminated.
e) Dirt or dust is visible in the field of
view.
f ) Visibility is poor.
· Image is not poor.
· Contrast is poor.
· Details are indistinct.
· Image glares.
28
Cause
Bulb is burned out.
Power cord is unplugged.
It has been turned OFF due to the
operation of the ECO mode.
Aperture and field iris diaphragms are
not opened wide enough.
Condenser is lowered too much.
Light path selector knob is set to the
position.
Light path selector knob is in an inter-
mediate position.
Revolving nosepiece is not correctly
engaged.
Condenser is not attached properly.
Revolving nosepiece is not attached
properly.
An objective that falls outside of the
condenser's illumination range is used.
Condenser is not properly centered.
Field iris diaphragm is stopped down
too far.
Bulb is not mounted correctly.
Dirt/dust on the eyepiece.
Dirt or the to surface of the condenser.
Dirt/dust on the specimen.
A non-UIS2/UIS objective is used.
Condenser is lowered too far.
Aperture iris diaphragm is stopped
down too far.
Revolving nosepiece is not mounted
properly.
Correction collar on correction collar
equipped objective is not properly
adjusted.
Front lens of objective is dirty.
Immersion oil is not being used with
an oil immersion objective.
Remedy
Replace the bulb. Note that, even
when the filament looks normal, there
may be a wire disconnection in the
terminal section.
Plug power cord into the power outlet.
Set the main switch to OFF then ON
again or press the ECO/NORMAL
switch twice.
Adjust them to proper sizes.
Adjust the condenser height position.
Move the knob to the
position.
Set the knob according to the obser-
vation method.
Make sure that the revolving nosepie-
ce clicks properly into place.
Re-attach it.
Push the side dovetail all the way until
it is stopped.
Use a condenser to match the pur-
pose.
Center the condenser.
Open the field iris diaphragm until an
optimum size.
Push the pins of halogen bulb all the
way until the stop position.
Clean thoroughly.
Use only UIS2/UIS series objectives
with this microscope.
Adjust the condenser height position.
Open the aperture iris diaphragm to
an optimum size.
Push the slide dovetail all the way until
it is stopped.
While focusing, turn the correction
collar to find the position with the best
contrast.
Clean objective.
Use immersion oil.