Olympus CK40-TBI Посібник з ремонту - Сторінка 6

Переглянути онлайн або завантажити pdf Посібник з ремонту для Мікроскоп Olympus CK40-TBI. Olympus CK40-TBI 39 сторінок. Tilting binocular tube

Olympus CK40-TBI Посібник з ремонту
CK40-TBI
B. INSPECTION
STANDARD
Part
Item
Standard
Method
BI unit
Difference
1.5 mm or less
Set the diopter at "0." Put a rigid Hat plate across the
CBinocular
between left and
left and right tube surface, insert a 1 mm thickness
tllbe
right tube
gauge between the flat plate and the tube surface. If
section)
length
the clearance is greater than 1 mm, the unit is out of
the specification.
Absolute
0.2 mm or less on the
parfocality
image surface
Combine the standard eyepiece KNO048 (with
Adapter-l), the focusing telescope @T-36), a
microscope stand (product), and the standard objective
(KNO016). Set the interpupillary distance at 62 mm,
read the parfocality difference between the reticle scale
of KNO048 (with Adapter-l) and the specimen of
KNO016 in right sleeve using the helicoid scale (1
graduation=O.l mm) of the KNO048.
TBIC unit
Optical axis
0.5 mm or less on the
In the observation state, observe a specimen whose
(Tilting
displacement by
image surface when the
center can be identified (ex. concentric circles) on the
section)
tilting
interpupillary distance is
stage using the KNO048 with Adapter-l.
Match the
62 mm
specimen center and the visual field center. While
tilting, read the displacement between' the specimen
center and the visual field center using the reticle scale
(1 graduation=O.l mm) of the KNO048 with Adapter-l.
Working force
up-Down400-9oog
Using a tension gauge, measure the working force after
(3920 mN - 8820 mN)
mounting the TBIC unit on the microscope stand
Down - Up 200 - 700 g
(CK40).
(1960 mN - 6860 mN)
B-2