CK40-TB I
B. INSPECTION
STANDARD
c-
1. Inspection
items and methods
(f- ?
P
Part
Item
Standard
Method
H unit
Interpupillary
50 mm to 76 mm
In the observation state, insert a thin sheet of paper
Binocular
distance
with graduations at the eye point position and measure
ube
adjustment
the interpupillary distance.
ection)
range
Interpupillary
1000 - 2000 g
Put a string on the sleeve and measure the working
distance
(9800 mN - 19600 mN)
force to actuate the interpupillary distance using a
working force
tension gauge.
Revolving axis
0.2 mm or less
Using the standard eyepiece (KNO048 with Adapter-l)
center
and a 4X or 10X objective, observe a specimen whose
center can be identified (ex. concentric circles) on the
stage.
Taking the right sleeve as the reference, match the
specimen center and the visual field center, open and
close the interpupillary distance to read the
displacement between the specimen center and the
visual field center in the right sleeve using the reticle
scale (1 graduation=O.l mm) of the KNO048.
Left/right
optical axis
On image surface:
Using the standard eyepiece (KNO048 with Adapter-l)
0.2 mm or less in vertical
and a 4X or 10X objective, observe a specimen whose
direction
center can be identified (ex. concentric circles) on the
0.2 mm or less in outward
stage.
direction
Taking the right sleeve as the reference, match the
0.4 mm or less in inward
specimen center and the visual field center and read
direction
the displacement between the specimen center and the
visual field center in the left sleeve using the reticle
scale (1 graduation=O.l mm) of the KNO048.
Absolute optical
0.1 mm or less on the
Combining the standard eyepiece (RN0048 with
axis (right
image surface
Adapter-l), a microscope stand (product), the standard
sleeve only)
objective (KNOO15), read the displacement between the
specimen center and the visual field center in the right
sleeve using the reticle scale (1 graduation=O.l mm) of
the RN0048 (with adapter-l).
Exit pupil
center
Within 20% of objective' s
Combining the centering telescope (KNOO29), a 4X or
exit pupil diameter
10X objective, and a microscope stand (product), read
the displacement between the exit pupil center of the
objective and the cross hairs center of centering
telescope.
Left/right
parfocality
difference
0.1 mm or less
Combine the standard eyepiece (KNOO48; with
Adapter-l), the focusing telescope @T-36), a
microscope stand (product), and the standard objective
(KNO016).
Operating the helicoid of the RN0048 (with Adapter-l)
separately in the left and right optical paths, focus on
the specimen in the RN0016 and read the parfocality
difference using the helicoid scale (1 graduation=O.l
mm) of the RN0048 (with Adapter-l).
B-i