2
Escaping the condenser top lens
During fluorescence observation, when the background light of the field of view becomes darker, the contrast of
TIP
the fluorescence observation image is improved. If the condenser turret is set to FL (fluorescence observation),
the background light of the field of view can be darker than that of other turret positions. If you want to make the
background light of the field of view darker effectively, escape the condenser top lens from the light path.
a
c
Using the finger
1
Insert the finger to the aperture of the stage, press the finger hook por-
a
tion
and move the top lens to the front (arrow direction).
Using the hook stick
If the finger cannot be inserted to the aperture of the stage
TIP
due to using the specimen holder for observing two slide
glasses or plain specimen holder, use the hook stick provided
with the B excitation fluorescence illuminator.
1
Insert the hook stick
hook it over the pin
the front (arrow direction).
Be sure to store the hook stick in a safe place.
NOTE
When using the plain specimen holder, it is recommended
TIP
b
to remove the top lens from the light path in advance prior to
observation.
b
in the insertion hole on the front of the stage,
c
near the top lens and move the top lens to
CX43-RFAB
9