Dini Argeo ALP Посібник з експлуатації - Сторінка 7
Переглянути онлайн або завантажити pdf Посібник з експлуатації для Ваги Dini Argeo ALP. Dini Argeo ALP 19 сторінок.
2.3 Level Balance
Level the balance on a firm, steady surface. Avoid locations with excessive air current, vibrations, heat
sources or rapid temperature changes.
2.4 Connect Power
For use with CSA Certified (or
equivalent approved) power
source, which must have a
limited circuit output.
2.5 Initial Calibration
Without InCal – Power on the balance by pressing
. Press and hold Print/Cal until [CAL] is
displayed. The display flashes the calibration mass needed. To select the alternate calibration weight
press No. Put the calibration mass on the pan. The display flashes [BUSY], then [CLEAr Pan]. Remove
the mass. When calibration is complete, [DONE] is displayed.
InCal – Press and hold Print/Cal until [CAL] is displayed. The display flashes [BUSY], when calibration
is complete, [DONE] is displayed.
Note: Calibrations should be performed after a warm up time of 60 minutes.
7