3. Operation
Count, APW Optimization, Percentage or specific units of measure must be activated in the MODE or
UNIT menus if they are not initially available.
3.1 Weighing Mode
Repeatedly press Unit until the desired unit icon is displayed.
Press
to zero the balance and then place objects to be weighed on the pan.
3.2 Count Mode
Use the Count mode to count parts of uniform weight.
To access Count Mode, press Unit until the display shows [Count].
Establish an Average Piece Weight (APW) – Each time a new type of part is to be counted, the nominal
weight of one piece (APW) must be established using a small quantity of pieces.
With [CLr.APwW] displayed, press No to use the previously saved APW, or press Yes to establish a new
APW. The display indicates the number of pieces to be used to establish the new APW. If a different
sample size is preferred, press No until the desired sample size is displayed (5, 10, 20, 50 or 100). Put the
specified number of pieces on the pan. Press Yes to accept new APW or No to abort.
Count – Place the quantity to be counted on the pan.
APW Optimization – Since the weight of each piece varies slightly, APW Optimization may be used to
increase the
accuracy of the count. The balance automatically recalculates the Average Piece Weight when the
number of parts on the pan is less than three times the original sample size. The display shows
[APwW.OPT] each time the APW is optimized.
3.3 Percent Mode
Use this mode to measure the weight of a sample as a percentage of a reference weight.
To access the Percent Mode, press Unit until the display shows [PErCEnt].
Establish a new Reference Weight - With Clear reference [CLr rEF] displayed, Press No to use the
previously saved Reference Weight. Press Yes to establish a new Reference Weight. Put the reference
sample on the pan and press Yes to accept or No to abort.
Percent – Place the object(s) to be compared to the reference weight on the pan.
8