Assembly for the triple-viewing model
1 Check that the serial numbers on the two components (on the main section and the
relay section) of the MTH relay unit are identical. (If the numbers are different,
observation will be adversely affected. In this case, please contact your dealer or
nearest Nikon representative.)
2 Adjust the height of the support stand so that the main observer section and the
secondary observer section are horizontal. This step is facilitated by first adjusting the
support stand to an approximate height before attaching the stand.
3 In the illustration below, the secondary observers are positioned on the right side of the
main observer, but they may also be positioned on the left. Attach the MTH main
splitter so that the pointer mounting surface is facing towards the front.
4 When mounting the attachment on the inverted microscopes TE300 and TE200, the TE-
TI intermediate eyepiece tube and E400/600 eyepiece tube are required. Attach the TE-
TI intermediate eyepiece tube instead of the original eyepiece tube and mount the MTH
main splitter on it.
See the illustration on P.12 for the installation of the TE-TI intermediate eyepiece tube.
ND filter (φ28 mm)
Clamp for 5
Ensure that the pointer
mounting surface faces
the front.
Clamp for 3
[Remarks]
If a trinocular eyepiece tube T is used for the main observer section, you can provide the image to the vertical
tube of the eyepiece tube by setting the optical path to the "BINO & PHOTO" position. If the ND filter is
removed in advance from the eyepiece tube mounting port of the MTH main splitter, the main and the secondary
observers will be able to view an image with more uniform brightness. This should be convenient such as when
performing the TV observation.
MTH main splitter
5
Clamp for 7
7
4
3
Main observer section
III. Assembly
MTH relay unit
1
Check that the
serial numbers
are the same.
MTH relay unit
Secondary
observer section
Place
horizontally
9
6
Height adjustment
clamp screw
Adjust to an
approximate height
and readjust later.
2
Main section
Upper surface
should also be
horizontal.
Clamp for 9
8
Clamp for 8
13