4
Place the imaging target in the visual field.
Set a specimen, direct the light path to the binocular part to
perform epi-fluorescence microscopy, and then put the
STORM imaging target into the visual field.
(For details on the procedure for epi-fluorescence
microscopy, see the user manual for the microscope.)
5
Configure camera settings.
In the [DU-897 Settings] control window (EM-CCD camera
settings), configure the settings as follows.
[Format For Live]:
[Format For Capture]: No Binning
Exposure time:
[Readout Mode]:
[EM Gain Multiplier]:
[Conversion Gain]:
[Desired Temperature]
(Commands -> Advanced
Camera Settings):
6
Wait until the temperature of the camera stabilizes at
about -70°C and [Desired temp. differs!] disappears.
It takes a few minutes for the temperature of the camera to
stabilize.
7
Put the objectives for STORM into the light path.
Objectives:
CFI Apo TIRF 100x oil (NA1.49) or
CFI Plan Apo VC 100x oil (NA1.40)
Use Nikon immersion oil Type B or Type NF for oil
immersion of the objectives.
When using CFI Apo TIRF 100x oil, adjust the correction
ring to suit the cover glass. If a cover glass No. 1-S
(No.1.5) is used, it is recommended that the position of the
correction ring be 0.165 mm.
Chapter 2
STORM Microscopy Operation
No Binning
Any setting (50 msec or 1
frame recommended)
EM Gain 10 MHz 14-bit
30
Maximum available value
-70°C
5
7