ALTONOVUS OR-01 Посібник із швидкого старту

Переглянути онлайн або завантажити pdf Посібник із швидкого старту для Вимірювальні прилади ALTONOVUS OR-01. ALTONOVUS OR-01 2 сторінки. Low compliance voltage ohmmeter

ALTONOVUS OR-01 Посібник із швидкого старту
DEVICE
USE AND CARE
Do Not immerse in,
or expose to liquid.
IP54 rated.
Do Not drop, expose to
shock, or place heavy
objects on unit.
Do Not apply pressure
to display.
Only use a clean,
dry cloth to clean
unit.
BEFORE USE
INSERT 2 x AA BATTERIES
2 x AA alkaline batteries will give in excess of 150
hours of continuous operation. Treat all batteries in
accordance with the manufacturer's instructions.
1. Remove battery compartment cover (H) applying
gentle pressure to finger pad.
2. Ensure clean contacts on batteries.
3. Insert new batteries, observing polarity
instructions in battery compartment.
4. Replace battery compartment cover.
If desired, remove the protective film from the front of
the ΩhmRanger-LCV. Peel carefully from below the
Positive
ΩhmRanger-LCV will still function normally if the film
is left in place.
SCREEN GUIDE
A
- The Resistance reading, displayed in MΩ, KΩ and Ω. In the example
above, 2.1MΩ.
B
- A small, oscillating Range Bar shows which range is currently in use.
An upper, longer Range Coverage Bar shows any movement across the
ranges during the reading (in this case, none). The box surrounding
shows the operating range.
C
- The Voltage being applied by ΩhmRanger-LCV at the
terminal, in this case 206.7mV, to measure the resistance.
D
- The active Range and the measured Current.
E
- The Signal to Noise Ratio (SNR). The SNR indicates how much noise,
variation, and AC components are in the signal, followed by Maximum and
Minimum Current samples during the reading. Higher resistances (lower
currents) are more susceptible to noise.
USE THIS QUICK START GUIDE IN CONJUNCTION WITH THE FULL INSTRUCTION MANUAL. DOWNLOAD AT:
Avoid extremes of
temperature. Max
accuracy in 20-25
degree Celsius range.
No User-serviceable
components. Do Not
open unit, as it voids
warranty and
calibration.
Minimum stimulus
applied is 206mV. Do
not expose unit to
currents over 800mA.
and Negative terminals.
Positive
MODEL: OR-01: LOW COMPLIANCE VOLTAGE OHMMETER.
QUICK START GUIDE (Rev 1.0 dated 10 MAR 22)
ALTONOVUS LTD. 37 OAST HOUSE CRESCENT, FARNHAM, SURREY,
GETTING STARTED - TAKE A READING
WARNING. Use
ΩhmRanger-LCV
with cables less
than 3 metres.
STEP 2: CONNECT LEADS
Connect the Resistance to be
measured between the
Positive
(RED) and Negative (BLACK)
terminals of the ΩhmRanger-LCV.
Leads (not supplied) should be
industry standard, uninsulated, 4mm
banana plug. Leads of shorter
length will help minimise any noise.
STEP 4: POWER OFF
After measurements
are taken, press the
POWER button
again to return
ΩhmRanger-LCV
to Standby mode.
A -
Positive
Terminal
B - Negative Terminal
C - Display Screen
D - On / Off Button
E - Menu Navigation
F - Select Button
G - Isolated USB Interface
H - Battery Compartment
STEP 1: POWER ON
After inserting batteries
as shown, press the
POWER button to start
ΩhmRanger-LCV.
STEP 3: TAKE MEASUREMENTS
Sampling at a high rate to
eliminate mains noise, Ωhm-
Ranger-LCV takes about 2.5
readings per second. It displays
the current in MΩ, KΩ and Ω. In
this example, 2.1MΩ.
A Range Bar below the digits
shows which ranges are, and have
been in use. The Range Bar
oscillates after each reading to
show how quickly readings are
occurring and to show that the
device is functioning correctly.
WWW.ALTONOVUS.COM