STERILIZATION AND MAINTENANCE:
Because of its simple design, the Aseptico ADU-17X Express II requires very little maintenance. Any maintenance that
is needed can be performed in minutes.
BLEEDING THE SYSTEM
If the unit will not be used for an extended period of time, or the unit might be subjected to freezing conditions, you
should bleed the system. Simply empty the contents of the water supply tank and install the tank back into the cap.
Operate the air/water syringe and handpiece with water coolant 'ON' until just air comes through the water line. Pack
unit and store as normal.
HANDPIECE FLUSH
Flush the handpiece for about 5 seconds after every patient, and about 20 seconds at the beginning of each day.
GENERAL CLEANING
The external surfaces of the case should be cleaned using a soft cloth moistened with a mild detergent solution. Any
external surfaces of the unit that are contacted during use should be wiped down with a soft cloth moistened with a
disinfectant at the beginning of each day and between each patient use.
WATER LINES
Disinfect the water lines weekly. Prepare a 1:10 bleach solution (1 part household bleach to 9 parts water). Remove
water reservoir and discard residual water. Replace empty water supply tank and air purge all waterlines. Fill water
supply tank with bleach solution. Run bleach solution through all lines. Allow bleach solution to stand in lines for 10
minutes. Remove water supply tank and discard bleach. Flush water supply tank and all lines thoroughly
with clean water. Air purge and leave lines dry until next clinical use.
VACUUM SYSTEM
The HVE and Saliva Ejector Valves are fully autoclavable. Remove the valves from the hoses before autoclaving. The
vacuum hoses should not be autoclaved. Clean hoses with a disinfectant solution. CAUTION: Use only NON-foaming
cleansers in the vaccum lines.
SYRINGE TIPS
The ADU-17X unit features a three-way air/water syringe with quick-
change autoclavable tips. To remove a tip, press on the locking
collar surrounding the tip socket and pull the used tip straight out of
the socket (see Fig. 8). To insert a new tip, press locking collar and
push tip into socket as far as it will go. Release collar and gently
tug on tip before using to ensure that tip is securely locked into
socket.
Syringe Tip Sterilization:
1) Remove contaminated syringe tip.
2) Remove all visible signs of contamination before autoclaving.
3) Autoclave tip at 132° C (270° F) for ten minutes.
4) Sterilize between each patient use.
NOTE: Since only the tips can be autoclaved, it is recommended that the Air/Water Syringe be bagged with a
disposable, single-use plastic sleeve between each patient use.
This device requires periodic maintenance to remove dust in and around the compressor every 12 months or if
accumulations are seen. The fan on the unit vents air from inside the device to the outside. With the unit operating,
use compressed air to spray into the intake vents and blow dust out of the unit.
8.
FIG. 8
DEPRESS
LOCKING COLLAR