OPERATION:
After you have set up the Express II unit and
have become familiar with its functions,
enable one of the following two Run Modes:
• Foot Switch Run Mode - Depressing
Foot Switch activates all components.
• Continuous Run Mode - Switch 'ON'
activates all components.
1. Connect the power cord to the unit and
plug into appropriate power source.
IMPORTANT: Ensure that the Voltage
Selector Insert is properly configured for
the power source - Refer to Page 4.
2. FOOT SWITCH RUN MODE:
a. Turn the Run Mode Switch on the front
of the Fan Housing to the "FOOT
SWITCH" position (See Fig. 6).
b. Depress the Foot
Switch: The unit will
pressurize and all
four system
components
(Handpiece, 3-Way
Syringe, HVE, and
Saliva Ejector) will
activate.
CAUTION:
Some new ADU-17X compressor motors
might require a 2-week "break-in" when first
used. As a precaution against stalling the
compressor or possibly blowing the fuses
during this break-in period, it is
recommended that either the handpiece or
the water bottle be removed from the
System for approximately 5 minutes before
each initial daily startup, during the first 2
weeks of use.
WARNING:
Always operate a high-speed handpiece with water
coolant. Operating a high-speed handpiece without
water coolant can cause thermal injury to the patient
and damage the handpiece. Consult the handpiece
manufacturer's instructions for additional information.
FIG. 6
RUN
MODE
SWITCH
3. CONTINUOUS RUN MODE:
a. Turn the Run Mode Switch on the front
of the Fan Housing to the
"CONTINUOUS" position.
b. The unit will pressurize and all four
system components (Handpiece, 3-Way
Syringe, HVE, and Saliva Ejector) will
activate. The Foot Switch is not
functional.
4. Turn Handpiece Switch On and adjust air
pressure and water coolant flow to the
desired settings (See Fig. 7). (Turn the
water coolant control knob clockwise to
decrease flow, or counterclockwise to
increase flow.)
FIG. 7 - HANDPIECE CONTROLS
HANDPIECE
ON/OFF
SWITCH
COOLANT
ON/OFF
SWITCH
AIR
CONTROL
VALVE
COOLANT
CONTROL
VALVE
CAVITRON
CONNECTOR
5. Use the 3-Way Air/Water Syringe as
necessary for irrigation or drying.
6. Adjust the vacuum pressure by moving
the valves on both the HVE and saliva
ejector vacuum heads.
7. Empty the Vacuum Waste Tank when it is
3/4 full.
7.